1stQ
B1AP00
Basis Z preloaded
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: clear | Preloaded: ✔ yes | Foldable: – |
Incision Width: 2.4 mm | Injector Size: – | hydrophilic |
Refractive Index: – | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: neutral | SA Correction: – | Toric: ✘ no |
Sphere: | 0 to 9 D (increment 1 D) 10 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | – |
Haigis: | a₀ = 1.243, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.46 |
Holladay 1: | SF = 1.67 |
SRK/T: | A = 118.9 |
Barrett: | – |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | – |
Hoffer Q: | – |
Holladay 1: | – |
SRK/T: | – |
Castrop: | – |
Cooke: | – |
We don't have enough (matching) clinical results, so we couldn't optimize constants for this lens.
1stQ
B1APY0
Basis Z preloaded gelb
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✔ yes | Foldable: – |
Incision Width: 2.4 mm | Injector Size: – | hydrophilic |
Refractive Index: – | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: neutral | SA Correction: – | Toric: ✘ no |
Sphere: | 0 to 9 D (increment 1 D) 10 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | – |
Haigis: | a₀ = 1.243, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.46 |
Holladay 1: | SF = 1.67 |
SRK/T: | A = 118.9 |
Barrett: | – |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | – |
Hoffer Q: | – |
Holladay 1: | – |
SRK/T: | – |
Castrop: | – |
Cooke: | – |
We don't have enough (matching) clinical results, so we couldn't optimize constants for this lens.
1stQ
B2EWYM
Basis Q progressiv (trifokal) gelb Add. +3.00
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✘ no | Foldable: ✔ yes |
Incision Width: 2.2 mm | Injector Size: – | hydrophilic |
Refractive Index: – | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 11 mm | Intended Location: capsular bag |
Optic Concept: multifocal | Haptic Design: Q-haptic | Optic Design: asphere |
Sph. Aberration: neutral | SA Correction: – | Toric: ✘ no |
Sphere: | –0 to 30 D (increment 0.5 D) 30 to 35 D (increment 1 D) |
Cylinder: | – |
Near Addition: | 3 D |
Intermediate Addition: | – |
Nominal: | – |
Haigis: | a₀ = 1.243, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.46 |
Holladay 1: | SF = 1.67 |
SRK/T: | A = 118.9 |
Barrett: | – |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | – |
Hoffer Q: | – |
Holladay 1: | – |
SRK/T: | – |
Castrop: | – |
Cooke: | – |
We don't have enough (matching) clinical results, so we couldn't optimize constants for this lens.
AJL Ophthalmic
LLY360
ORIOL MAX
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✘ no | Foldable: ✔ yes |
Incision Width: 2.75 mm | Injector Size: – | hydrophilic |
Refractive Index: 1.462 | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 10.75 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: PLATE | Optic Design: sphere |
Sph. Aberration: – | SA Correction: – | Toric: ✘ no |
Sphere: | −5 to 12 D (increment 1 D) 12.5 to 25 D (increment 0.5 D) 26 to 40 D (increment 1 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 117.3 |
Haigis: | a₀ = 0.95, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 4.56 |
Holladay 1: | SF = 1.37 |
SRK/T: | A = 118 |
Barrett: | – |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | – |
Hoffer Q: | – |
Holladay 1: | – |
SRK/T: | – |
Castrop: | – |
Cooke: | – |
We don't have enough (matching) clinical results, so we couldn't optimize constants for this lens.
AJL Ophthalmic
LLY6002
ORIOL
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✘ no | Foldable: ✔ yes |
Incision Width: 2.2 mm | Injector Size: – | hydrophilic |
Refractive Index: 1.462 | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 10.75 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: PLATE | Optic Design: sphere |
Sph. Aberration: – | SA Correction: – | Toric: ✘ no |
Sphere: | −5 to 12 D (increment 1 D) 12.5 to 25 D (increment 0.5 D) 26 to 40 D (increment 1 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 117.3 |
Haigis: | a₀ = 0.95, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 4.56 |
Holladay 1: | SF = 1.37 |
SRK/T: | A = 118 |
Barrett: | – |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | – |
Hoffer Q: | – |
Holladay 1: | – |
SRK/T: | – |
Castrop: | – |
Cooke: | – |
We don't have enough (matching) clinical results, so we couldn't optimize constants for this lens.
Alcon
AcrySof AU00T0
UltraSert
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: – | hydrophobic |
Refractive Index: 1.55 | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: STABLEFORCE | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.2 µm | Toric: ✘ no |
Sphere: | 6 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.7 (ultrasound) |
Haigis: | a₀ = −0.769, a₁ = 0.234, a₂ = 0.217 |
Hoffer Q: | pACD = 5.64 |
Holladay 1: | SF = 1.84 |
SRK/T: | A = 119 |
Barrett: | DF = 5 |
Olsen: | – |
Provided by manufacturer or distributor and adopted from the ULIB platform. Optimization based on 5363 clinical results.
Haigis: | a₀ = −0.025, a₁ = 0.31, a₂ = 0.173 (Mean abs. pred. error: 0.3253 D) |
Hoffer Q: | pACD = 5.57 (Mean abs. pred. error: 0.3336 D) |
Holladay 1: | SF = 1.8 (Mean abs. pred. error: 0.3174 D) |
SRK/T: | A = 118.97 (Mean abs. pred. error: 0.345 D) |
Castrop: | C = 0.35912, H = 0.07948, R = −0.18353 (Mean abs. pred. error: 0.3794 D) |
Cooke: | A = 118.96478 (Mean abs. pred. error: 0.30841 D) |
Calculated by IOL Con from 535 clinical results.
Ethnicity: 7 % caucasian, 93 % unknown
Biometer: 30 % Zeiss IOLMaster, 25 % Haag-Streit Lenstar, 45 % unknown
Alcon
AcrySof SN6CWS
AcrySert
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: – | hydrophobic |
Refractive Index: 1.55 | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: STABLEFORCE | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.2 µm | Toric: ✘ no |
Sphere: | 6 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.7 (ultrasound) |
Haigis: | a₀ = −0.769, a₁ = 0.234, a₂ = 0.217 |
Hoffer Q: | pACD = 5.64 |
Holladay 1: | SF = 1.84 |
SRK/T: | A = 119 |
Barrett: | DF = 5 |
Olsen: | – |
Provided by manufacturer or distributor and adopted from the ULIB platform. Optimization based on 5363 clinical results.
Haigis: | a₀ = −0.025, a₁ = 0.31, a₂ = 0.173 (Mean abs. pred. error: 0.3253 D) |
Hoffer Q: | pACD = 5.57 (Mean abs. pred. error: 0.3336 D) |
Holladay 1: | SF = 1.8 (Mean abs. pred. error: 0.3174 D) |
SRK/T: | A = 118.97 (Mean abs. pred. error: 0.345 D) |
Castrop: | C = 0.35912, H = 0.07948, R = −0.18353 (Mean abs. pred. error: 0.3794 D) |
Cooke: | A = 118.96478 (Mean abs. pred. error: 0.30841 D) |
Calculated by IOL Con from 535 clinical results.
Ethnicity: 7 % caucasian, 93 % unknown
Biometer: 30 % Zeiss IOLMaster, 25 % Haag-Streit Lenstar, 45 % unknown
Alcon
Clareon CNA0T0
Clareon AutonoMe
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: 2.2 mm | Injector Size: 2 mm | hydrophobic |
Refractive Index: 1.55 | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: STABLEFORCE | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.2 µm | Toric: ✘ no |
Sphere: | 6 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.8 (ultrasound) |
Haigis: | – |
Hoffer Q: | – |
Holladay 1: | SF = 1.824 |
SRK/T: | A = 119.1 |
Barrett: | DF = 5 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = −0.399, a₁ = 0.209, a₂ = 0.211 (Mean abs. pred. error: 0.324 D) |
Hoffer Q: | pACD = 5.8 (Mean abs. pred. error: 0.3303 D) |
Holladay 1: | SF = 2.01 (Mean abs. pred. error: 0.3159 D) |
SRK/T: | A = 119.33 (Mean abs. pred. error: 0.325 D) |
Castrop: | – |
Cooke: | A = 119.32638 (Mean abs. pred. error: 0.3027 D) |
Calculated by IOL Con from 740 clinical results.
Ethnicity: 76 % caucasian, 24 % unknown
Biometer: 85 % Zeiss IOLMaster, 10 % Haag-Streit Lenstar, 5 % unknown
Alcon
Clareon SY60WF
Clareon
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✘ no | Foldable: ✔ yes |
Incision Width: – | Injector Size: – | hydrophobic |
Refractive Index: 1.55 | Abbe Number: – | Achromatic: – |
Optic Ø: – | Haptic Ø: – | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: STABLEFORCE | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.2 µm | Toric: ✘ no |
Sphere: | 6 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.8 (ultrasound) |
Haigis: | – |
Hoffer Q: | – |
Holladay 1: | SF = 1.824 |
SRK/T: | A = 119.1 |
Barrett: | DF = 5 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = −0.399, a₁ = 0.209, a₂ = 0.211 (Mean abs. pred. error: 0.324 D) |
Hoffer Q: | pACD = 5.8 (Mean abs. pred. error: 0.3303 D) |
Holladay 1: | SF = 2.01 (Mean abs. pred. error: 0.3159 D) |
SRK/T: | A = 119.33 (Mean abs. pred. error: 0.325 D) |
Castrop: | – |
Cooke: | A = 119.32638 (Mean abs. pred. error: 0.3027 D) |
Calculated by IOL Con from 740 clinical results.
Ethnicity: 76 % caucasian, 24 % unknown
Biometer: 85 % Zeiss IOLMaster, 10 % Haag-Streit Lenstar, 5 % unknown
Bausch + Lomb
Eye-CEE
EYECPRE
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✘ no | Optic Material: acrylic | Haptic Material: PMMA |
Filter: clear | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: 2.8 mm | Injector Size: – | hydrophobic |
Refractive Index: 1.52 | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 12.5 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: C loop | Optic Design: sphere |
Sph. Aberration: – | SA Correction: – | Toric: ✘ no |
Sphere: | 10 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.8 |
Haigis: | a₀ = 1.55, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.74 |
Holladay 1: | SF = 1.95 |
SRK/T: | A = 119.2 |
Barrett: | – |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | – |
Hoffer Q: | – |
Holladay 1: | – |
SRK/T: | – |
Castrop: | – |
Cooke: | – |
We don't have enough (matching) clinical results, so we couldn't optimize constants for this lens.
Bausch + Lomb
EyeCee One
EYEC1
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✘ no | Foldable: ✔ yes |
Incision Width: 2.2 mm | Injector Size: – | hydrophobic |
Refractive Index: 1.52 | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.14 µm | Toric: ✘ no |
Sphere: | 1 to 30 D (increment 1 D) 10 to 27 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.4 |
Haigis: | a₀ = 1.675, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 6 |
Holladay 1: | SF = 2.13 |
SRK/T: | A = 119.7 |
Barrett: | LF = 2.25, DF = 0 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = 1.81505 (with fixed a₁ = 0.4, a₂ = 0.1) (Mean abs. pred. error: 0.3839 D) |
Hoffer Q: | pACD = 5.90391 (Mean abs. pred. error: 0.3732 D) |
Holladay 1: | SF = 2.14328 (Mean abs. pred. error: 0.3743 D) |
SRK/T: | A = 119.60273 (Mean abs. pred. error: 0.386 D) |
Castrop: | – |
Cooke: | – |
Calculated by IOL Con from 50 clinical results.
Ethnicity: 100 % caucasian
Biometer: 100 % Zeiss IOLMaster
Bausch + Lomb
EyeCee One CRYSTAL
EYEC1CRY
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: clear | Preloaded: ✘ no | Foldable: ✔ yes |
Incision Width: 2.2 mm | Injector Size: – | hydrophobic |
Refractive Index: 1.52 | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.14 µm | Toric: ✘ no |
Sphere: | 1 to 30 D (increment 1 D) 10 to 27 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.4 |
Haigis: | a₀ = 1.675, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 6 |
Holladay 1: | SF = 2.13 |
SRK/T: | A = 119.7 |
Barrett: | LF = 2.25, DF = 0 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = 1.81505 (with fixed a₁ = 0.4, a₂ = 0.1) (Mean abs. pred. error: 0.3839 D) |
Hoffer Q: | pACD = 5.90391 (Mean abs. pred. error: 0.3732 D) |
Holladay 1: | SF = 2.14328 (Mean abs. pred. error: 0.3743 D) |
SRK/T: | A = 119.60273 (Mean abs. pred. error: 0.386 D) |
Castrop: | – |
Cooke: | – |
Calculated by IOL Con from 50 clinical results.
Ethnicity: 100 % caucasian
Biometer: 100 % Zeiss IOLMaster
Bausch + Lomb
Quatrix Evolutive
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: clear | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: 2.8 mm | Injector Size: – | hydrophilic |
Refractive Index: 1.465 | Abbe Number: – | Achromatic: – |
Optic Ø: 6.14 mm | Haptic Ø: 10.8 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: four-point-haptic | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.1 µm | Toric: ✘ no |
Sphere: | 10 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 119.6 |
Haigis: | a₀ = 1.91, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 6.04 |
Holladay 1: | SF = 2.28 |
SRK/T: | A = 119.8 |
Barrett: | – |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | – |
Hoffer Q: | – |
Holladay 1: | – |
SRK/T: | – |
Castrop: | – |
Cooke: | – |
We don't have enough (matching) clinical results, so we couldn't optimize constants for this lens.
Bausch + Lomb
SofPort AO
LI61AOR
Sulcus fixation possible
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✘ no | Optic Material: silicone | Haptic Material: PMMA |
Filter: clear | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: 2.2 mm | Injector Size: – | hydrophobic |
Refractive Index: 1.43 | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: neutral | SA Correction: – | Toric: ✘ no |
Sphere: | 0 to 34 D (increment 1 D) 5 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118 |
Haigis: | a₀ = 0.057, a₁ = 0.186, a₂ = 0.171 |
Hoffer Q: | pACD = 5.29 |
Holladay 1: | SF = 1.51 |
SRK/T: | A = 118.5 |
Barrett: | LF = 1.66, DF = 0 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | – |
Hoffer Q: | – |
Holladay 1: | – |
SRK/T: | – |
Castrop: | – |
Cooke: | – |
We don't have enough (matching) clinical results, so we couldn't optimize constants for this lens.
Bausch + Lomb
Versario Classic
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: clear | Preloaded: ✘ no | Foldable: ✔ yes |
Incision Width: 2.2 mm | Injector Size: – | hydrophilic |
Refractive Index: 1.465 | Abbe Number: – | Achromatic: – |
Optic Ø: 6.5 mm | Haptic Ø: 12.5 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.1 µm | Toric: ✘ no |
Sphere: | −10 to 40 D (increment 1 D) 10 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 120 |
Haigis: | a₀ = 1.8, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.96 |
Holladay 1: | SF = 2.2 |
SRK/T: | A = 119.7 |
Barrett: | – |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = 1.86258 (with fixed a₁ = 0.4, a₂ = 0.1) (Mean abs. pred. error: 0.3378 D) |
Hoffer Q: | pACD = 6.00008 (Mean abs. pred. error: 0.3455 D) |
Holladay 1: | SF = 2.26008 (Mean abs. pred. error: 0.3607 D) |
SRK/T: | A = 119.7829 (Mean abs. pred. error: 0.3743 D) |
Castrop: | – |
Cooke: | – |
Calculated by IOL Con from 54 clinical results.
Ethnicity: 100 % unknown
Biometer: 100 % Haag-Streit Lenstar
Biotech Europe Meditech
Optiflex AXIS TMFA6
Hydrophobic Aspheric Natural Yellow Toric IOL
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✘ no | Foldable: ✔ yes |
Incision Width: 2.8 mm | Injector Size: 1.975 mm | hydrophobic |
Refractive Index: 1.48 | Abbe Number: 49 | Achromatic: ✔ yes |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.2 µm | Toric: ✔ yes |
Equivalent: | 10 to 30 D (increment 0.5 D) |
Cylinder: | 1 to 1.5 D (increment 0.5 D) 1.5 to 6 D (increment 0.75 D) |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.5 |
Haigis: | a₀ = 1.243, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.53 |
Holladay 1: | SF = 1.76 |
SRK/T: | A = 118.9 |
Barrett: | LF = 1.83 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = −1.08432, a₁ = 0.30567, a₂ = 0.21794 (Mean abs. pred. error: 0.3197 D) |
Hoffer Q: | pACD = 5.52977 (Mean abs. pred. error: 0.3276 D) |
Holladay 1: | SF = 1.74001 (Mean abs. pred. error: 0.2949 D) |
SRK/T: | A = 118.82113 (Mean abs. pred. error: 0.2882 D) |
Castrop: | – |
Cooke: | – |
Calculated by IOL Con from 199 clinical results.
Ethnicity: 100 % unknown
Biometer: 100 % unknown
Biotech Europe Meditech
Optiflex GENE MPA6
Hydrophilic Aspheric Monofocal IOL
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: clear | Preloaded: ✘ no | Foldable: ✔ yes |
Incision Width: 2.8 mm | Injector Size: 1.975 mm | hydrophilic |
Refractive Index: 1.462 | Abbe Number: 42 | Achromatic: ✔ yes |
Optic Ø: 6 mm | Haptic Ø: 12.5 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: neutral | SA Correction: 0 µm | Toric: ✘ no |
Sphere: | 5 to 15 D (increment 1 D) 15 to 25 D (increment 0.5 D) 25 to 30 D (increment 1 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118 |
Haigis: | a₀ = 1.03, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.13 |
Holladay 1: | SF = 1.37 |
SRK/T: | A = 118.3 |
Barrett: | LF = 1.52 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = 0.65315, a₁ = 0.30847, a₂ = 0.12324 (Mean abs. pred. error: 0.4372 D) |
Hoffer Q: | pACD = 5.10189 (Mean abs. pred. error: 0.4385 D) |
Holladay 1: | SF = 1.3315 (Mean abs. pred. error: 0.4271 D) |
SRK/T: | A = 118.17938 (Mean abs. pred. error: 0.4111 D) |
Castrop: | – |
Cooke: | – |
Calculated by IOL Con from 198 clinical results.
Ethnicity: 100 % unknown
Biometer: 100 % unknown
Biotech Europe Meditech
Optiflex GENESIS COMFORT LMFAC6
Preloaded Hydrophobic Aspheric Monofocal IOL
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: clear | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: 2.2 mm | Injector Size: 1.925 mm | hydrophobic |
Refractive Index: 1.524 | Abbe Number: 47 | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.2 µm | Toric: ✘ no |
Sphere: | 5 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.5 |
Haigis: | a₀ = 1.03, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.25 |
Holladay 1: | SF = 1.46 |
SRK/T: | A = 118.4 |
Barrett: | – |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = 1.06865 (with fixed a₁ = 0.4, a₂ = 0.1) (Mean abs. pred. error: 0.8988 D) |
Hoffer Q: | pACD = 5.29793 (Mean abs. pred. error: 0.8387 D) |
Holladay 1: | SF = 1.51782 (Mean abs. pred. error: 0.8065 D) |
SRK/T: | A = 118.4935 (Mean abs. pred. error: 0.7591 D) |
Castrop: | – |
Cooke: | – |
Calculated by IOL Con from 98 clinical results.
Ethnicity: 100 % unknown
Biometer: 100 % Zeiss IOLMaster
Biotech Europe Meditech
Optiflex MPS6
Hydrophilic Monofocal IOL
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: clear | Preloaded: ✘ no | Foldable: ✔ yes |
Incision Width: 2.8 mm | Injector Size: 1.975 mm | hydrophilic |
Refractive Index: 1.462 | Abbe Number: 42 | Achromatic: ✔ yes |
Optic Ø: 6 mm | Haptic Ø: 12.5 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: C loop | Optic Design: sphere |
Sph. Aberration: – | SA Correction: – | Toric: ✘ no |
Sphere: | −10 to 1 D (increment 1 D) 1 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118 |
Haigis: | a₀ = 0.94, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.19 |
Holladay 1: | SF = 1.42 |
SRK/T: | A = 118.4 |
Barrett: | LF = 1.57 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = −0.44257, a₁ = 0.13261, a₂ = 0.19875 (Mean abs. pred. error: 0.446 D) |
Hoffer Q: | pACD = 5.15748 (Mean abs. pred. error: 0.4206 D) |
Holladay 1: | SF = 1.39912 (Mean abs. pred. error: 0.3969 D) |
SRK/T: | A = 118.29931 (Mean abs. pred. error: 0.386 D) |
Castrop: | – |
Cooke: | – |
Calculated by IOL Con from 195 clinical results.
Ethnicity: 100 % unknown
Biometer: 100 % unknown
Biotech Europe Meditech
Optiflex OPUS AXIS TDFA6
Hydrophobic Aspheric Natural Yellow Multifocal Toric IOL
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✘ no | Foldable: ✔ yes |
Incision Width: 2.8 mm | Injector Size: 1.975 mm | hydrophobic |
Refractive Index: 1.48 | Abbe Number: 49 | Achromatic: ✔ yes |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: multifocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.2 µm | Toric: ✔ yes |
Equivalent: | 10 to 30 D (increment 0.5 D) |
Cylinder: | 1 to 1.5 D (increment 0.5 D) 1.5 to 3 D (increment 0.75 D) |
Near Addition: | 3 D |
Intermediate Addition: | – |
Nominal: | A = 118.5 |
Haigis: | a₀ = 1.309, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.52 |
Holladay 1: | SF = 1.74 |
SRK/T: | A = 119 |
Barrett: | LF = 1.88 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = 0.97111 (with fixed a₁ = 0.4, a₂ = 0.1) (Mean abs. pred. error: 0.3218 D) |
Hoffer Q: | pACD = 5.18135 (Mean abs. pred. error: 0.3048 D) |
Holladay 1: | SF = 1.4221 (Mean abs. pred. error: 0.2577 D) |
SRK/T: | A = 118.32121 (Mean abs. pred. error: 0.2546 D) |
Castrop: | – |
Cooke: | – |
Calculated by IOL Con from 72 clinical results.
Ethnicity: 100 % unknown
Biometer: 100 % Zeiss IOLMaster
Biotech Europe Meditech
Optiflex OPUS D3FA6
Hydrophobic Aspheric Natural Yellow Multifocal IOL
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✘ no | Foldable: ✔ yes |
Incision Width: 2.8 mm | Injector Size: 1.975 mm | hydrophobic |
Refractive Index: 1.48 | Abbe Number: 49 | Achromatic: ✔ yes |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: multifocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.2 µm | Toric: ✘ no |
Sphere: | 7.5 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | 3 D |
Intermediate Addition: | – |
Nominal: | A = 118.5 |
Haigis: | a₀ = 0.879, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.17 |
Holladay 1: | SF = 1.45 |
SRK/T: | A = 118.35 |
Barrett: | LF = 1.54 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = −0.74364, a₁ = 0.3098, a₂ = 0.19055 (Mean abs. pred. error: 0.3935 D) |
Hoffer Q: | pACD = 5.23747 (Mean abs. pred. error: 0.3911 D) |
Holladay 1: | SF = 1.45754 (Mean abs. pred. error: 0.361 D) |
SRK/T: | A = 118.41137 (Mean abs. pred. error: 0.3654 D) |
Castrop: | – |
Cooke: | – |
Calculated by IOL Con from 178 clinical results.
Ethnicity: 100 % unknown
Biometer: 100 % unknown
Biotech Vision Care
EYECRYL PLUS NATURAL HD ASHFY6002
Hydrophobic Aspheric Natural Yellow Monofocal IOL
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✘ no | Foldable: ✔ yes |
Incision Width: 2.2 mm | Injector Size: 1.925 mm | hydrophobic |
Refractive Index: 1.524 | Abbe Number: 47 | Achromatic: ✔ yes |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.2 µm | Toric: ✘ no |
Sphere: | 5 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.5 |
Haigis: | a₀ = 1.077, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.3 |
Holladay 1: | SF = 1.51 |
SRK/T: | A = 118.65 |
Barrett: | – |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = 1.7923 (with fixed a₁ = 0.4, a₂ = 0.1) (Mean abs. pred. error: 0.8649 D) |
Hoffer Q: | pACD = 5.96282 (Mean abs. pred. error: 0.8503 D) |
Holladay 1: | SF = 2.18533 (Mean abs. pred. error: 0.8137 D) |
SRK/T: | A = 119.4884 (Mean abs. pred. error: 0.7829 D) |
Castrop: | – |
Cooke: | – |
Calculated by IOL Con from 60 clinical results.
Ethnicity: 100 % unknown
Biometer: 20 % Zeiss IOLMaster, 80 % unknown
Biotech Vision Care
EYECRYL SERT PAHFY600F
Preloaded Hydrophobic Aspheric Natural Yellow Monofocal IOL
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: 2.2 mm | Injector Size: 1.925 mm | hydrophobic |
Refractive Index: 1.524 | Abbe Number: 47 | Achromatic: ✔ yes |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.2 µm | Toric: ✘ no |
Sphere: | 5 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.5 |
Haigis: | a₀ = 1.03, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.25 |
Holladay 1: | SF = 1.46 |
SRK/T: | A = 118.4 |
Barrett: | – |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = −0.00831, a₁ = 0.21509, a₂ = 0.16911 (Mean abs. pred. error: 0.5058 D) |
Hoffer Q: | pACD = 5.29091 (Mean abs. pred. error: 0.4806 D) |
Holladay 1: | SF = 1.50943 (Mean abs. pred. error: 0.4534 D) |
SRK/T: | A = 118.49281 (Mean abs. pred. error: 0.4334 D) |
Castrop: | – |
Cooke: | – |
Calculated by IOL Con from 206 clinical results.
Ethnicity: 100 % unknown
Biometer: 100 % Zeiss IOLMaster
Biotech Vision Care
EYECRYL SERT PLHF2
Preloaded Hydrophobic Aspheric Monofocal IOL
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: clear | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: 2.2 mm | Injector Size: 1.925 mm | hydrophobic |
Refractive Index: 1.524 | Abbe Number: 47 | Achromatic: ✔ yes |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.2 µm | Toric: ✘ no |
Sphere: | 5 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.5 |
Haigis: | a₀ = 1.03, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.25 |
Holladay 1: | SF = 1.46 |
SRK/T: | A = 118.4 |
Barrett: | – |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = 1.06865 (with fixed a₁ = 0.4, a₂ = 0.1) (Mean abs. pred. error: 0.8988 D) |
Hoffer Q: | pACD = 5.29793 (Mean abs. pred. error: 0.8387 D) |
Holladay 1: | SF = 1.51782 (Mean abs. pred. error: 0.8065 D) |
SRK/T: | A = 118.4935 (Mean abs. pred. error: 0.7591 D) |
Castrop: | – |
Cooke: | – |
Calculated by IOL Con from 98 clinical results.
Ethnicity: 100 % unknown
Biometer: 100 % Zeiss IOLMaster
HOYA
XC1
Vivinex iSert
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: clear | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: 1.7 mm | hydrophobic |
Refractive Index: – | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.18 µm | Toric: ✘ no |
Sphere: | 6 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.9 |
Haigis: | a₀ = −1.04585, a₁ = 0.25472, a₂ = 0.2291 |
Hoffer Q: | pACD = 5.70046 |
Holladay 1: | SF = 1.92783 |
SRK/T: | A = 119.193 |
Barrett: | LF = 1.98 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = −1.04585, a₁ = 0.25472, a₂ = 0.2291 (Mean abs. pred. error: 0.3335 D) |
Hoffer Q: | pACD = 5.70046 (Mean abs. pred. error: 0.3557 D) |
Holladay 1: | SF = 1.92783 (Mean abs. pred. error: 0.3456 D) |
SRK/T: | A = 119.19313 (Mean abs. pred. error: 0.3624 D) |
Castrop: | C = 0.32488, H = 0.12666, R = 0.15478 (Mean abs. pred. error: 0.3472 D) |
Cooke: | A = 119.2307 (Mean abs. pred. error: 0.3391 D) |
Calculated by IOL Con from 2857 clinical results.
Ethnicity: 31 % caucasian, 7 % japanese, 61 % unknown
Biometer: 29 % Zeiss IOLMaster, 4 % Allegro, 1 % Heidelberg E. Anterion, 0 % Haag-Streit Lenstar, 66 % unknown
HOYA
XC1-SP
Vivinex multiSert
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: clear | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: 1.7 mm | hydrophobic |
Refractive Index: – | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.18 µm | Toric: ✘ no |
Sphere: | 6 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.9 |
Haigis: | a₀ = −1.04585, a₁ = 0.25472, a₂ = 0.2291 |
Hoffer Q: | pACD = 5.70046 |
Holladay 1: | SF = 1.92783 |
SRK/T: | A = 119.193 |
Barrett: | LF = 1.98 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = −1.04585, a₁ = 0.25472, a₂ = 0.2291 (Mean abs. pred. error: 0.3335 D) |
Hoffer Q: | pACD = 5.70046 (Mean abs. pred. error: 0.3557 D) |
Holladay 1: | SF = 1.92783 (Mean abs. pred. error: 0.3456 D) |
SRK/T: | A = 119.19313 (Mean abs. pred. error: 0.3624 D) |
Castrop: | C = 0.32488, H = 0.12666, R = 0.15478 (Mean abs. pred. error: 0.3472 D) |
Cooke: | A = 119.2307 (Mean abs. pred. error: 0.3391 D) |
Calculated by IOL Con from 2857 clinical results.
Ethnicity: 31 % caucasian, 7 % japanese, 61 % unknown
Biometer: 29 % Zeiss IOLMaster, 4 % Allegro, 1 % Heidelberg E. Anterion, 0 % Haag-Streit Lenstar, 66 % unknown
HOYA
XY1
Vivinex iSert
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: 1.7 mm | hydrophobic |
Refractive Index: – | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.18 µm | Toric: ✘ no |
Sphere: | 6 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.9 |
Haigis: | a₀ = −1.04585, a₁ = 0.25472, a₂ = 0.2291 |
Hoffer Q: | pACD = 5.70046 |
Holladay 1: | SF = 1.92783 |
SRK/T: | A = 119.193 |
Barrett: | LF = 1.98 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = −1.04585, a₁ = 0.25472, a₂ = 0.2291 (Mean abs. pred. error: 0.3335 D) |
Hoffer Q: | pACD = 5.70046 (Mean abs. pred. error: 0.3557 D) |
Holladay 1: | SF = 1.92783 (Mean abs. pred. error: 0.3456 D) |
SRK/T: | A = 119.19313 (Mean abs. pred. error: 0.3624 D) |
Castrop: | C = 0.32488, H = 0.12666, R = 0.15478 (Mean abs. pred. error: 0.3472 D) |
Cooke: | A = 119.2307 (Mean abs. pred. error: 0.3391 D) |
Calculated by IOL Con from 2857 clinical results.
Ethnicity: 31 % caucasian, 7 % japanese, 61 % unknown
Biometer: 29 % Zeiss IOLMaster, 4 % Allegro, 1 % Heidelberg E. Anterion, 0 % Haag-Streit Lenstar, 66 % unknown
HOYA
XY1-GT / XY1-GPT
Vivinex Gemetric Toric / Vivinex Gemetric Plus Toric preloaded in multiSert
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: 1.7 mm | hydrophobic |
Refractive Index: – | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: multifocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.18 µm | Toric: ✔ yes |
Equivalent: | 10 to 30 D (increment 0.5 D) |
Cylinder: (t-notation) | T2: 1 D T3: 1.5 D T4: 2.25 D T5: 3 D T6: 3.75 D |
Near Addition: | 3.5 D |
Intermediate Addition: | 1.75 D |
Nominal: | A = 119 |
Haigis: | a₀ = −0.0199, a₁ = 0.3437, a₂ = 0.162 |
Hoffer Q: | pACD = 5.548 |
Holladay 1: | SF = 1.789 |
SRK/T: | A = 118.998 |
Barrett: | LF = 1.88 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = −0.01987, a₁ = 0.34366, a₂ = 0.16204 (Mean abs. pred. error: 0.3695 D) |
Hoffer Q: | pACD = 5.54795 (Mean abs. pred. error: 0.3645 D) |
Holladay 1: | SF = 1.78891 (Mean abs. pred. error: 0.3562 D) |
SRK/T: | A = 118.99752 (Mean abs. pred. error: 0.3712 D) |
Castrop: | – |
Cooke: | A = 118.88245 (Mean abs. pred. error: 0.35076 D) |
Calculated by IOL Con from 913 clinical results.
Ethnicity: 100 % unknown
Biometer: 98 % Zeiss IOLMaster, 1 % Pentacam, 0 % Nidek AL-Scan, 0 % Haag-Streit Lenstar, 0 % Ziemer Galilei
HOYA
XY1-G / XY1-GP
Vivinex Gemetric / Vivinex Gemetric Plus preloaded in multiSert
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: 1.7 mm | hydrophobic |
Refractive Index: – | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: multifocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.18 µm | Toric: ✘ no |
Sphere: | 10 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | 3.5 D |
Intermediate Addition: | 1.75 D |
Nominal: | A = 119 |
Haigis: | a₀ = −0.0199, a₁ = 0.3437, a₂ = 0.162 |
Hoffer Q: | pACD = 5.548 |
Holladay 1: | SF = 1.789 |
SRK/T: | A = 118.998 |
Barrett: | LF = 1.88 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = −0.01987, a₁ = 0.34366, a₂ = 0.16204 (Mean abs. pred. error: 0.3695 D) |
Hoffer Q: | pACD = 5.54795 (Mean abs. pred. error: 0.3645 D) |
Holladay 1: | SF = 1.78891 (Mean abs. pred. error: 0.3562 D) |
SRK/T: | A = 118.99752 (Mean abs. pred. error: 0.3712 D) |
Castrop: | – |
Cooke: | A = 118.88245 (Mean abs. pred. error: 0.35076 D) |
Calculated by IOL Con from 913 clinical results.
Ethnicity: 100 % unknown
Biometer: 98 % Zeiss IOLMaster, 1 % Pentacam, 0 % Nidek AL-Scan, 0 % Haag-Streit Lenstar, 0 % Ziemer Galilei
HOYA
XY1-SP
Vivinex multiSert
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: 1.7 mm | hydrophobic |
Refractive Index: – | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.18 µm | Toric: ✘ no |
Sphere: | 6 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.9 |
Haigis: | a₀ = −1.04585, a₁ = 0.25472, a₂ = 0.2291 |
Hoffer Q: | pACD = 5.70046 |
Holladay 1: | SF = 1.92783 |
SRK/T: | A = 119.193 |
Barrett: | LF = 1.98 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = −1.04585, a₁ = 0.25472, a₂ = 0.2291 (Mean abs. pred. error: 0.3335 D) |
Hoffer Q: | pACD = 5.70046 (Mean abs. pred. error: 0.3557 D) |
Holladay 1: | SF = 1.92783 (Mean abs. pred. error: 0.3456 D) |
SRK/T: | A = 119.19313 (Mean abs. pred. error: 0.3624 D) |
Castrop: | C = 0.32488, H = 0.12666, R = 0.15478 (Mean abs. pred. error: 0.3472 D) |
Cooke: | A = 119.2307 (Mean abs. pred. error: 0.3391 D) |
Calculated by IOL Con from 2857 clinical results.
Ethnicity: 31 % caucasian, 7 % japanese, 61 % unknown
Biometer: 29 % Zeiss IOLMaster, 4 % Allegro, 1 % Heidelberg E. Anterion, 0 % Haag-Streit Lenstar, 66 % unknown
HumanOptics
ASPIRA-aXA
XL-Optik
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: clear | Preloaded: ✘ no | Foldable: ✔ yes |
Incision Width: – | Injector Size: – | hydrophilic |
Refractive Index: – | Abbe Number: – | Achromatic: – |
Optic Ø: 7 mm | Haptic Ø: 11 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: plate with cut-out | Optic Design: asphere |
Sph. Aberration: neutral | SA Correction: – | Toric: ✘ no |
Sphere: | 10 to 30 D (increment 0.5 D) −10 to 9 D (increment 1 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.3 (optical) |
Haigis: | a₀ = 1.667, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.89 |
Holladay 1: | SF = 2.13 |
SRK/T: | A = 119.5 |
Barrett: | LF = 2.18 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = −0.52091, a₁ = 0.26934, a₂ = 0.21524 (Mean abs. pred. error: 0.3614 D) |
Hoffer Q: | pACD = 5.92412 (Mean abs. pred. error: 0.3737 D) |
Holladay 1: | SF = 2.16633 (Mean abs. pred. error: 0.3503 D) |
SRK/T: | A = 119.56808 (Mean abs. pred. error: 0.3512 D) |
Castrop: | – |
Cooke: | A = 119.63669 (Mean abs. pred. error: 0.33932 D) |
Calculated by IOL Con from 220 clinical results.
Ethnicity: 74 % caucasian, 26 % unknown
Biometer: 74 % Zeiss IOLMaster, 26 % unknown
HumanOptics
ASPIRA-aXA Safeloader
XL-Optik
Preloaded Safeloader autoloading container
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: clear | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: – | hydrophilic |
Refractive Index: – | Abbe Number: – | Achromatic: – |
Optic Ø: 7 mm | Haptic Ø: 11 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: plate with cut-out | Optic Design: asphere |
Sph. Aberration: neutral | SA Correction: – | Toric: ✘ no |
Sphere: | 10 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.3 (optical) |
Haigis: | a₀ = 1.667, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.89 |
Holladay 1: | SF = 2.13 |
SRK/T: | A = 119.5 |
Barrett: | LF = 2.18 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = −0.52091, a₁ = 0.26934, a₂ = 0.21524 (Mean abs. pred. error: 0.3614 D) |
Hoffer Q: | pACD = 5.92412 (Mean abs. pred. error: 0.3737 D) |
Holladay 1: | SF = 2.16633 (Mean abs. pred. error: 0.3503 D) |
SRK/T: | A = 119.56808 (Mean abs. pred. error: 0.3512 D) |
Castrop: | – |
Cooke: | A = 119.63669 (Mean abs. pred. error: 0.33932 D) |
Calculated by IOL Con from 220 clinical results.
Ethnicity: 74 % caucasian, 26 % unknown
Biometer: 74 % Zeiss IOLMaster, 26 % unknown
HumanOptics
ASPIRA-aXAY Safeloader
XL-Optik
Preloaded Safeloader autoloading container
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yes | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: – | hydrophilic |
Refractive Index: – | Abbe Number: – | Achromatic: – |
Optic Ø: 7 mm | Haptic Ø: 11 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: palte with cut-out | Optic Design: asphere |
Sph. Aberration: neutral | SA Correction: – | Toric: – |
Sphere: | 10 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 118.3 (optical) |
Haigis: | a₀ = 1.667, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.89 |
Holladay 1: | SF = 2.13 |
SRK/T: | A = 119.5 |
Barrett: | LF = 2.18 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = −0.52091, a₁ = 0.26934, a₂ = 0.21524 (Mean abs. pred. error: 0.3614 D) |
Hoffer Q: | pACD = 5.92412 (Mean abs. pred. error: 0.3737 D) |
Holladay 1: | SF = 2.16633 (Mean abs. pred. error: 0.3503 D) |
SRK/T: | A = 119.56808 (Mean abs. pred. error: 0.3512 D) |
Castrop: | – |
Cooke: | A = 119.63669 (Mean abs. pred. error: 0.33932 D) |
Calculated by IOL Con from 220 clinical results.
Ethnicity: 74 % caucasian, 26 % unknown
Biometer: 74 % Zeiss IOLMaster, 26 % unknown
HumanOptics
DIFF-aA Safeloader
Preloaded Safeloader autoloading container
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: clear | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: – | hydrophilic |
Refractive Index: – | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 12.5 mm | Intended Location: capsular bag |
Optic Concept: multifocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: neutral | SA Correction: – | Toric: ✘ no |
Sphere: | 10 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | 3.5 D |
Intermediate Addition: | – |
Nominal: | A = 118.4 |
Haigis: | a₀ = 1.28, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.5 |
Holladay 1: | SF = 1.73 |
SRK/T: | A = 118.9 |
Barrett: | LF = 1.83 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | – |
Hoffer Q: | – |
Holladay 1: | – |
SRK/T: | – |
Castrop: | – |
Cooke: | – |
We don't have enough (matching) clinical results, so we couldn't optimize constants for this lens.
HumanOptics
DIFF-aAY Safeloader
Preloaded Safeloader autoloading container
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: – | hydrophilic |
Refractive Index: – | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 12.5 mm | Intended Location: capsular bag |
Optic Concept: multifocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: neutral | SA Correction: – | Toric: ✘ no |
Sphere: | 10 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | 3.5 D |
Intermediate Addition: | – |
Nominal: | A = 118.4 |
Haigis: | a₀ = 1.28, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.5 |
Holladay 1: | SF = 1.73 |
SRK/T: | A = 118.9 |
Barrett: | LF = 1.83 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | – |
Hoffer Q: | – |
Holladay 1: | – |
SRK/T: | – |
Castrop: | – |
Cooke: | – |
We don't have enough (matching) clinical results, so we couldn't optimize constants for this lens.
HumanOptics
TORICADIFF-aA Safeloader
Preloaded Safeloader autoloading container
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: clear | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: – | hydrophilic |
Refractive Index: – | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 12.5 mm | Intended Location: capsular bag |
Optic Concept: multifocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: neutral | SA Correction: – | Toric: ✔ yes |
Sphere: | 10 to 30 D (increment 0.5 D) |
Cylinder: | 1 to 6 D (increment 0.5 D) |
Near Addition: | 3.5 D |
Intermediate Addition: | – |
Nominal: | A = 118.4 (optical) |
Haigis: | a₀ = 0.885, a₁ = 0.312, a₂ = 0.125 |
Hoffer Q: | pACD = 5.36 |
Holladay 1: | SF = 1.6 |
SRK/T: | A = 118.7 |
Barrett: | LF = 1.73 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | – |
Hoffer Q: | – |
Holladay 1: | – |
SRK/T: | – |
Castrop: | – |
Cooke: | – |
We don't have enough (matching) clinical results, so we couldn't optimize constants for this lens.
HumanOptics
TORICADIFF-aAY
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✘ no | Foldable: ✔ yes |
Incision Width: – | Injector Size: – | hydrophilic |
Refractive Index: – | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 12.5 mm | Intended Location: capsular bag |
Optic Concept: multifocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: neutral | SA Correction: – | Toric: ✔ yes |
Sphere: | 10 to 30 D (increment 0.5 D) |
Cylinder: | 1 to 6 D (increment 0.5 D) |
Near Addition: | 3.5 D |
Intermediate Addition: | – |
Nominal: | A = 118.4 (optical) |
Haigis: | a₀ = 0.885, a₁ = 0.312, a₂ = 0.125 |
Hoffer Q: | pACD = 5.36 |
Holladay 1: | SF = 1.6 |
SRK/T: | A = 118.7 |
Barrett: | LF = 1.73 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | – |
Hoffer Q: | – |
Holladay 1: | – |
SRK/T: | – |
Castrop: | – |
Cooke: | – |
We don't have enough (matching) clinical results, so we couldn't optimize constants for this lens.
HumanOptics
TORICADIFF-aAY Safeloader
Preloaded Safeloader autoloading container
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: – | hydrophilic |
Refractive Index: – | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 12.5 mm | Intended Location: capsular bag |
Optic Concept: multifocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: neutral | SA Correction: – | Toric: ✔ yes |
Sphere: | 10 to 30 D (increment 0.5 D) |
Cylinder: | 1 to 6 D (increment 0.5 D) |
Near Addition: | 3.5 D |
Intermediate Addition: | – |
Nominal: | A = 118.4 (optical) |
Haigis: | a₀ = 0.885, a₁ = 0.312, a₂ = 0.125 |
Hoffer Q: | pACD = 5.36 |
Holladay 1: | SF = 1.6 |
SRK/T: | A = 118.7 |
Barrett: | LF = 1.73 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | – |
Hoffer Q: | – |
Holladay 1: | – |
SRK/T: | – |
Castrop: | – |
Cooke: | – |
We don't have enough (matching) clinical results, so we couldn't optimize constants for this lens.
HumanOptics
TRIVA-aA Safeloader
trifocal,preloaded Safeloader®(SL) autoloading container
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: no | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: – | hydrophilic |
Refractive Index: – | Abbe Number: 56 | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 12.5 mm | Intended Location: – |
Optic Concept: multifocal | Haptic Design: – | Optic Design: asphere |
Sph. Aberration: neutral | SA Correction: – | Toric: ✔ yes |
Equivalent: | 10 to 30 D (increment 0.5 D) |
Cylinder: | 1 to 6 D (increment 0.5 D) |
Near Addition: | 3.5 D |
Intermediate Addition: | 1.75 D |
Nominal: | A = 118.4 (optical) |
Haigis: | a₀ = 1.561, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.813 |
Holladay 1: | SF = 2.113 |
SRK/T: | A = 119.554 |
Barrett: | LF = 1.99 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = 0.63546, a₁ = 0.27699, a₂ = 0.15267 (Mean abs. pred. error: 0.3252 D) |
Hoffer Q: | pACD = 5.68923 (Mean abs. pred. error: 0.3283 D) |
Holladay 1: | SF = 1.96215 (Mean abs. pred. error: 0.3289 D) |
SRK/T: | A = 119.29305 (Mean abs. pred. error: 0.3434 D) |
Castrop: | – |
Cooke: | – |
Calculated by IOL Con from 105 clinical results.
Ethnicity: 37 % caucasian, 63 % unknown
Biometer: 100 % Zeiss IOLMaster
HumanOptics
TRIVA-aAY Safeloader
trifocal, blue light filter, preloaded Safeloader® (SL) autoloading container
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: – | hydrophilic |
Refractive Index: – | Abbe Number: 56 | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 12.5 mm | Intended Location: – |
Optic Concept: multifocal | Haptic Design: – | Optic Design: asphere |
Sph. Aberration: neutral | SA Correction: – | Toric: – |
Sphere: | 10 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | 3.5 D |
Intermediate Addition: | 1.75 D |
Nominal: | A = 118.4 (optical) |
Haigis: | a₀ = 1.561, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.813 |
Holladay 1: | SF = 2.113 |
SRK/T: | A = 119.554 |
Barrett: | LF = 1.99 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | a₀ = 0.63546, a₁ = 0.27699, a₂ = 0.15267 (Mean abs. pred. error: 0.3252 D) |
Hoffer Q: | pACD = 5.68923 (Mean abs. pred. error: 0.3283 D) |
Holladay 1: | SF = 1.96215 (Mean abs. pred. error: 0.3289 D) |
SRK/T: | A = 119.29305 (Mean abs. pred. error: 0.3434 D) |
Castrop: | – |
Cooke: | – |
Calculated by IOL Con from 105 clinical results.
Ethnicity: 37 % caucasian, 63 % unknown
Biometer: 100 % Zeiss IOLMaster
HumanOptics
TRIVAT-aA Safeloader
trifocal,toric,preloaded Safeloader®(SL) autoloading container
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: no | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: – | hydrophilic |
Refractive Index: – | Abbe Number: 56 | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 12.5 mm | Intended Location: capsular bag |
Optic Concept: multifocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: neutral | SA Correction: 0 µm | Toric: ✔ yes |
Equivalent: | 10 to 30 D (increment 0.5 D) |
Cylinder: | 1 to 6 D (increment 0.5 D) |
Near Addition: | 3.5 D |
Intermediate Addition: | 1.75 D |
Nominal: | A = 118.9 (optical) |
Haigis: | a₀ = 1.426, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.63 |
Holladay 1: | SF = 1.9 |
SRK/T: | A = 119.21 |
Barrett: | LF = 1.99 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | – |
Hoffer Q: | – |
Holladay 1: | – |
SRK/T: | – |
Castrop: | – |
Cooke: | – |
We don't have enough (matching) clinical results, so we couldn't optimize constants for this lens.
HumanOptics
TRIVAT-aAY Safeloader
trifocal,toric,blue light filter,preloaded Safeloader®(SL) autoloading containe
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: yellow | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: – | hydrophilic |
Refractive Index: – | Abbe Number: 56 | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 12.5 mm | Intended Location: capsular bag |
Optic Concept: multifocal | Haptic Design: C loop | Optic Design: asphere |
Sph. Aberration: neutral | SA Correction: 0 µm | Toric: ✔ yes |
Equivalent: | 10 to 30 D (increment 0.5 D) |
Cylinder: | 1 to 6 D (increment 0.5 D) |
Near Addition: | 3.5 D |
Intermediate Addition: | 1.75 D |
Nominal: | A = 118.9 (optical) |
Haigis: | a₀ = 1.426, a₁ = 0.4, a₂ = 0.1 |
Hoffer Q: | pACD = 5.63 |
Holladay 1: | SF = 1.9 |
SRK/T: | A = 119.21 |
Barrett: | LF = 1.99 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | – |
Hoffer Q: | – |
Holladay 1: | – |
SRK/T: | – |
Castrop: | – |
Cooke: | – |
We don't have enough (matching) clinical results, so we couldn't optimize constants for this lens.
Johnson and Johnson Vision
AAB00
SENSAR 1PC
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: UV | Preloaded: ✘ no | Foldable: ✔ yes |
Incision Width: – | Injector Size: – | hydrophobic |
Refractive Index: 1.47 | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: Haptics offset from optic | Optic Design: sphere |
Sph. Aberration: – | SA Correction: – | Toric: ✘ no |
Sphere: | 6 to 30 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 119 (optical) |
Haigis: | a₀ = −1.004, a₁ = 0.182, a₂ = 0.232 |
Hoffer Q: | pACD = 5.56 |
Holladay 1: | SF = 1.78 |
SRK/T: | A = 119 |
Barrett: | LF = 1.88, DF = 5.5 |
Olsen: | – |
Provided by manufacturer or distributor and adopted from the ULIB platform. Optimization based on 415 clinical results.
Haigis: | a₀ = −0.42313, a₁ = 0.39715, a₂ = 0.1796 (Mean abs. pred. error: 0.3574 D) |
Hoffer Q: | pACD = 5.57118 (Mean abs. pred. error: 0.3914 D) |
Holladay 1: | SF = 1.79701 (Mean abs. pred. error: 0.3608 D) |
SRK/T: | A = 118.97399 (Mean abs. pred. error: 0.3633 D) |
Castrop: | C = 0.32572, H = 0.04246, R = −0.14461 (Mean abs. pred. error: 0.3813 D) |
Cooke: | A = 118.8056 (Mean abs. pred. error: 0.3596 D) |
Calculated by IOL Con from 1609 clinical results.
Ethnicity: 86 % caucasian, 14 % unknown
Biometer: 91 % Zeiss IOLMaster, 9 % unknown
Johnson and Johnson Vision
DCB00
Tecnis 1 Piece ZCB00 with Tecnis Simplicity Delivery System
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: UV | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: – | hydrophobic |
Refractive Index: 1.47 | Abbe Number: – | Achromatic: – |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: monofocal | Haptic Design: Haptics offset from optic | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.27 µm | Toric: ✘ no |
Sphere: | 5 to 34 D (increment 0.5 D) |
Cylinder: | – |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 119.3 (optical) |
Haigis: | a₀ = −1.302, a₁ = 0.21, a₂ = 0.251 |
Hoffer Q: | pACD = 5.8 |
Holladay 1: | SF = 2.02 |
SRK/T: | A = 119.3 |
Barrett: | LF = 2.04, DF = 4.5 |
Olsen: | – |
Provided by manufacturer or distributor and adopted from the ULIB platform. Optimization based on 2041 clinical results.
Haigis: | a₀ = −1.35616, a₁ = 0.25919, a₂ = 0.24884 (Mean abs. pred. error: 0.3016 D) |
Hoffer Q: | pACD = 5.84477 (Mean abs. pred. error: 0.3354 D) |
Holladay 1: | SF = 2.06197 (Mean abs. pred. error: 0.3292 D) |
SRK/T: | A = 119.39951 (Mean abs. pred. error: 0.3594 D) |
Castrop: | C = 0.34655, H = 0.04406, R = 0.18895 (Mean abs. pred. error: 0.33625 D) |
Cooke: | A = 119.44614 (Mean abs. pred. error: 0.28737 D) |
Calculated by IOL Con from 693 clinical results.
Ethnicity: 100 % unknown
Biometer: 12 % Haag-Streit Lenstar, 9 % Zeiss IOLMaster, 79 % unknown
Johnson and Johnson Vision
DFW (FDA approved)
Tecnis Synergy Toric II Optiblue with Tecnis Simplicity Delivery System (FDA)
This intraocular lens is currently not available from the manufacturer / distributor.
Single Piece: ✔ yes | Optic Material: acrylic | Haptic Material: – |
Filter: UV. Violet blocking | Preloaded: ✔ yes | Foldable: ✔ yes |
Incision Width: – | Injector Size: – | hydrophobic |
Refractive Index: 1.47 | Abbe Number: – | Achromatic: ✔ yes |
Optic Ø: 6 mm | Haptic Ø: 13 mm | Intended Location: capsular bag |
Optic Concept: EDoF | Haptic Design: New squared and frosted haptic design | Optic Design: asphere |
Sph. Aberration: correcting | SA Correction: −0.27 µm | Toric: ✔ yes |
Equivalent: | 5 to 34 D (increment 0.5 D) |
Cylinder: (t-notation) | DFW150: 1.5 D DFW225: 2.25 D DFW300: 3 D DFW375: 3.75 D |
Near Addition: | – |
Intermediate Addition: | – |
Nominal: | A = 119.3 (optical) |
Haigis: | – |
Hoffer Q: | pACD = 5.7 |
Holladay 1: | SF = 1.96 |
SRK/T: | A = 119.3 |
Barrett: | LF = 2.04, DF = 4.5 |
Olsen: | – |
Provided by manufacturer or distributor.
Haigis: | – |
Hoffer Q: | – |
Holladay 1: | – |
SRK/T: | – |
Castrop: | – |
Cooke: | – |
We don't have enough (matching) clinical results, so we couldn't optimize constants for this lens.